OM-550 u46u
CALIBRATION PROCEDURE
Using the following Calibration procedure we recommend that the OM-550 is
recalibrated every 6 months.
Setup:
1. Perform a Manual Reset
2. Press the PROG key to enter the programming mode
3. Expect a printout similar to the following:
Figure A-2
4. Answer the prompts as follows:
Figure A-3
PRESS OK
PRESS 000005
PRESS NO
PRESS (CURRENT DATE)
PRESS (CURRENT TIME)
PRESS NO
PRESS OK
PRESS YES
PROGRAMMING IS DONE